Horiba HE-960CW Conductivity meter

NTN TECHNOLOGY CO.,LTD

Contact Information:

Description
  • Versatile measurement range
  • 2-channel simultaneous measurement
    2 sensors can simultaneously measure and compare conductivity, etc. at 2 locations
  • 4 transmission outputs
    Equipped with
  • Equipped with RS-485 communication function
  • Equipped with USP <645> water quality judgment function
    Automatically determines conductivity standards for pharmaceutical water such as pyrogen-free water required by USP <645>
  • Temperature monitoring during steam sterilization (SIP) is also possible because it can display up to 160°C
    (*Varies depending on the combined sensor)

Specification

MODEL HE-960CW
SENSOR INPUT 2 channels
MEASUREMENT METHOD 2-electrode method
CELL CONSTANT 0.01/cm, 0.1/cm or 1.0/cm
MEASUREMENT RANGE cell constant (/cm) 0.01 0.1 1.0
Conductivity range μS/cm 2.000/20.00 2.000* / 20.00 / 200.0 / 2000* 200.0/2000
mS/m 0.2000/2.000 0.2000* / 2.000 / 20.00 / 200.0* 20.00/200.0
TDS conversion mg/L 2.000/20.00 20.00 / 200.0 200.0/2000
* Applicable only to sanitary sensor (FS-01 series)
temperature 0 to 100°C (Selectable display of 1 or 2 digits after the decimal point)
Salt removal rate 0.0 to 100.0%
Conductivity difference According to the measurement range of sensor 2 (CH 2 )
REPEATABILITY Within ±0.5% full scale (with equivalent input)
, but TDS conversion display within ±1.5% full scale
2000 μS/cm, 200.0 mS/m range within ±5% full scale
LINEARITY Within ±0.5% full scale (with equivalent input)
, but TDS conversion display within
±1.5% full scale
DISPLAY/OUTPUT RANGE conductivity 110% of measurement range
temperature -30 to 160℃
Salt removal rate 0-100%
Conductivity difference 110% of sensor 2 (CH 2 ) measurement range
TRANSMISSION OUTPUT Number of output points: 4 points

  • 4 to 20mA DC/0 to 20mA DC input/output isolated type
  • Maximum load resistance: 900Ω
  • Transmission output range: Can be set arbitrarily within the measurement range .

 

CONTACT OUTPUT Number of output points: 4 points

  • Contact form: Relay contact R1 to R3 SPST (1a) R4 SPDT (1c)
  • Contact capacity: 240V AC 3A, 30V DC 3A (resistive load)
  • Contact function: Upper and lower limit operation (ON/OFF control) delay time, hysteresis
  • Output contents: Selectable setting from each measurement target, USP judgment, error alarm, maintenance
    (However, R1 and 2, R3 and 4 are common contact)